Viability using Acridine Orange/Propidium Iodide

Introduction to Cell Viability Measurement

Since the quality of the cell sample is vital for potential downstream experiments, viability measurements are routinely performed in many laboratories. Choosing the correct method for conducting cell viability measurement is essential for obtaining consistent accurate results. Here we review a common method for measuring cell viability: Acridine Orange/Propidium Iodide.

Ask a Specialist
Request a Quote
TB vs. Fluorescent

Using Acridine Orange/Propidium Iodide (AO/PI) to Measure Cell Viability

Acridine orange (AO) and propidium iodide (PI) are nucleic acid binding dyes that can be used to measure the cell viability. Both AO and PI are considered skin irritants and may be harmful if ingested or inhaled; they are currently not classified as carcinogenic. Since AO is cell permeable, all stained nucleated cells generate a green fluorescence. PI (~668 Daltons) only enters cells with compromised membranes and therefore dying, dead, and necrotic nucleated cells stained with PI generate a red fluorescence. When cells are stained with both AO and PI, live nucleated cells only fluoresce green and dead nucleated cells only fluoresce red. This is due to Förster resonance energy transfer (FRET); where the PI signal absorbs the AO signal producing no spill-over or double positive results. Additionally, other membrane-exclusion viability dyes such as: ethidium bromide (EB), 7AAD, SYTOX green/red, DRAQ5 and others may also be used instead of PI. Cell viability is calculated by examining the ratio of the number of live to the number of dead fluorescing cells. This assay can not only be used to measure the viability of nucleated cells in cell culture and purified samples but also in complicated samples such as PBMC, whole blood, bone marrow, bronchoalveolar lavage, tumor digests, primary samples and many more.

Bright field image of a peripheral blood sample

Bright field image of a peripheral blood sample

Live nucleated cells stained with Acridine Orange

Live nucleated cells stained with Acridine Orange

Dead nucleated cells stained with Propidium Iodide

Dead nucleated cells stained with Propidium Iodide

AO/PI Protocol: Sample Preparation and Analysis

Simple, User-friendly Cellometer Procedure

With the Cellometer Vision, Auto 2000, and K2 just 20µl of sample is added to the Cellometer Counting Chamber. Imaging and analysis of the samples is completed in less than 30 seconds. Bright field and fluorescent cell images can be viewed to check cell morphology and verify cell counting. Total cell count, concentration, and mean diameter are automatically displayed.

AO/PI Protocol

  1. Obtain Nexcelom AO/PI solution: View All Reagents
  2. Stain cell sample at 1:1 with AO/PI solution
  3. Load counting chamber slide and analyze
Pipette 20 µl of sample into a disposable slide

1. Pipette 20 µl of sample into a disposable slide

Insert slide into the cell counting instrument

2. Insert slide into the instrument

Select assay from a drop down menu

3. Select assay from a drop down menu

Cell Counting Data

4. Click count, acquire image and view cell count, concentration, diameter

Viability Measurement for Messy Samples Using AO/PI

Tumor Digest Samples

Tumor digest samples are collected from the patient at the primary source of the tumor. Solid tumors are first digested and processed before they are utilized in experiments. These are inherently messy samples. They may contain tissue fragments and cellular debris making it difficult to identify and count cells of interest. With AO/PI only the nucleated cells are stained and analyzed.

Tumor Digest Images

Tumor Digest Images

Leuko Pak Images

Leuko Pak Images

Cord Blood Images

Cord Blood Images

Click edit button to change this text.Both Leuko Pak and Cord blood samples contain a large amount of red blood cells, platelets, cell debris making it impossible to count in bright field. Using AO/PI the concentration and viability of mononuclear cells can be measured.

Viability Measurement in PBMCs

The quality of peripheral blood mononuclear cell (PBMC) samples after isolation, during experimentation, as well as before and after cryopreservation needs to be assessed. The two micrographs (right) show live mononuclear cells stained with acridine orange in green, and dead cells stained with PI in red.

Learn more about PBMC »

PBMC Images

PBMC image

Conclusions & Cellometer Selection Guide

The Cellometer line of instruments can quickly and accurately obtain the viability of the cell sample by trypan blue or fluorescent methods. Additionally, the system automatically reports an accurate cell count, concentration, and cell size in a single 20 µl assay. The fluorescent based Cellometer instruments (Auto 2000, Vision, and K2) also allow researcher to count the total number of nucleated cells in a fresh clinical sample without the need to lyse red blood cells. The table below outlines the recommended Cellometer instrument and assay use for each selected application.

Cellometer Mini Cellometer Auto T4 Cellometer Auto 1000 Cellometer Auto 2000 Cellometer Vision / Vision CBA
Purified cells– no debris TB TB TB TB or AOPI TB or AOPI
Isolated MNCs – without lysing RBCs NR NR NR AOPI AOPI
Fresh BM, CB, WB, MNCs w/ lysing of RBCs
TB TB TB TB or AOPI TB or AOPI
Fresh BM, CB, WB without lysing NR NR NR AOPI AOPI
Frozen BM, CB, WB TB TB TB TB or AOPI TB or AOPI
Digested tumors and BAL with tissue/cellular debris NR NR NR AOPI AOPI

AOPI = Acridine Orange/Prodidium Iodide, TB = Trypan Blue, NR = not recommended, BM = Bone Marrow, CB = Cord Blood, WB = Whole Blood, MNCs = Mononuclear cells