Cell Concentration & Viability for Clinical Samples

Compatible with Cellometer:

Cellometer Dual-Fluorescence Counting Method

For viability determination, 20µl of sample is mixed with 20µl of Cellometer AO/PI Staining Solution. The acridine orange (AO) dye stains DNA in all nucleated cells, generating green fluorescence. Propidium iodide (PI) stains DNA in all cells with compromised cell membranes, generating red fluorescence. In cells stained with both AO and PI, the green fluorescence is absorbed by the red fluorescence via FRET (fluorescence resonance energy transfer). All dead nucleated cells fluoresce red (image below right). All live nucleated cells fluoresce green (image below left). Live and dead cells are counted in the fluorescent images.

Red blood cells (RBCs), platelets, and debris do not fluoresce and are excluded from cell counts, so sample lysis is not required with the AO/PI viability method.

Cellometer Dual-Fluorescence Counting Method

Back to top ↑

TNC Viability for Fresh Clinical Samples

TNC Viability

Because fresh clinical samples contain a very high percentage of platelets and red blood cells, incoming samples are sometimes not evaluated for TNC concentration. In some labs, samples are lysed for TNC analysis, but cell viability cannot be determined.

The Cellometer AO/PI dual-fluorescence method can be used to easily measure both TNC concentration and viability in fresh clinical samples, such as bone marrow, cord blood, and peripheral blood without lysing.

AO/PI Analysis Procedure:

  • Dilute bone marrow, cord blood or peripheral blood sample 1:10 with PBS.
  • Combine 20µl of diluted sample with 20µl of AO/PI Staining Solution.
  • Add 20µl of stained sample to the Cellometer Counting Chamber and insert into the Auto 2000 or Vision Image Cytometry System.
  • The Cellometer System automatically captures and analyzes bright field and fluorescent images.
  • The automated data report provides cell count, concentration, and mean diameter for both live and dead nucleated cells along with the % viability for the sample.

Fresh Bone Marrow Aspirate Analysis Results

Bright Field Image
(All cells) Fresh bone marrow aspirate brightfeild cell image
Green Fluorescent Image
(Live nucleated cells) Fresh bone marrow aspirate green fluorescent cell image
Red Fluorescent Image
(Dead nucleated cells) Fresh bone marrow aspirate red fluorescent cell image
Data Report Fresh Bone Marrow Aspirate Analysis Results

Fresh Cord Blood Analysis Results

Bright Field Image
(All cells) Fresh Cord Blood brightfeild image
Green Fluorescent Image
(Live nucleated cells) Fresh Cord Blood Green Fluorescent Image
Red Fluorescent Image
(Dead nucleated cells) Fresh Cord Blood red Fluorescent Image
Data Report Fresh Cord Blood results

Fresh Peripheral Blood Analysis Results

Bright Field Image
(All cells) Fresh Peripheral Blood brightfeild image
Green Fluorescent Image
(Live nucleated cells) Fresh Peripheral Blood Green Fluorescent Image
Red Fluorescent Image
(Dead nucleated cells) Fresh Peripheral Blood red Fluorescent Image
Data Report Fresh Peripheral Blood results

The report at left is automatically displayed on the screen.

A printed report with images can be generated directly from the Cellometer software.

Back to top ↑

Analysis of MNC and PBMC

Following removal of RBCs and platelets via density gradient separation, samples are often counted manually using Trypan Blue staining to calculate viability. Because samples can contain varying amounts of RBC contamination following density gradient separation, using the AO/PI dual-fluorescence method for samples at this stage offers improved accuracy and reproducibility.

AO/PI Analysis Procedure:

  1. Combine 20µl of sample with 20µl of AO/PI Staining Solution.
  2. Add 20µl of stained sample to the Cellometer Counting Chamber and insert into the Auto 2000 or Vision Image Cytometry System.
  3. The Cellometer System automatically captures and analyzes bright field and fluorescent images.
  4. The automated data report provides cell count, concentration, and mean diameter for both live and dead nucleated cells along with the % viability for the sample.

PBMC Sample Analysis Results

Bright Field Image
(All cells) PBMC Sample brightfeild image
Green Fluorescent Counted
Image
(Live nucleated cells) PBMC Sample Green Fluorescent Image
Red Fluorescent Counted
Image
(Dead nucleated cells) PBMC Sample red Fluorescent Image
Data Report PBMC Sample Analysis Results
As demonstrated below, there can be RBC contamination in samples following density gradient separation.

Using the AO/PI method eliminates potential error caused by counting of RBCs.

AO/PI method eliminates potential error caused by counting of RBCs

RBCs present in the PBMC sample are visible in the bright field image acquired with a Cellometer Vision 10x instrument (far left).

Because RBCs do not contain DNA, they are not stained by AO and are not counted as live nucleated cells in the green fluorescent image (near left).

Back to top ↑

Impact of RBC Contamination

Impact of RBC Contamination

Red blood cells look like white blood cells under the microscope. They are included in manual TNC counts.

Platelets, WBCs, and RBCs can be identified and counted in images taken with the Cellometer Vision instrument to measure the amount of RBC contamination in a sample.

Platelets, WBCs, and RBCs

RBC Contamination is Present Following Density Gradient Separation

RBC Contamination


In 15 random PBMC samples evaluated, 4 samples showed a high degree of RBC contamination and 6 samples showed moderate RBC contamination.

RBC contamination is present in MNC and PBMC samples.

The amount of RBC contamination varies from one sample to another.

Manual Counts Overestimate PBMC Counts & Viability in Samples with RBCs

Manual Counts Overestimate PBMC Counts


For samples containing high RBC contamination, manual counts over-estimated WBC concentration by ~25 to 50%.

AO/PI cell counts closely matched actual PBMC counts.

AO/PI provided a more accurate PBMC concentration and viability in samples with RBC contamination.

Back to top ↑

Analysis of Isolated Cell Populations from Purified Clinical Samples

Following immune-magnetic separation, samples display a high degree of purity and can be counted accurately using a wide range of counting methods, including both Trypan Blue and AO/PI. Whether isolated from bone marrow, cord blood, or peripheral blood, isolated cell populations are typically very clean, with no RBC or platelet contamination. Several examples of specific cell populations isolated from different types of samples are shown below. They are all very similar in appearance. Click on the CD34+ videos at the top right of this page for a demonstration of Cellometer AO/PI analysis of isolated cell populations.

CD56+ NK Cells from NPB

CD56+ NK Cells from NPB

CD34+ Stem Cells from CB

CD34+ Stem Cells from CB

CD19+ B Cells from NPB

CD19+ B Cells from NPB

Back to top ↑

AO/PI Correlation to Manual Counting for Clinical Samples

As part of a recent collaboration with AllCells, a provider of quality primary cells, Nexcelom scientists compared two popular bright field counting methods to Cellometer AO/PI dual-fluorescence counting. Results correlated well for both the Methylene Blue counting method (for all sample types) and the Trypan Blue counting method (for isolated cell populations).

Strong correlation to manual counting enables many labs to implement the change to the Cellometer AO/PI counting method with minimal documentation and testing.

AO/PI and Methylene Blue TNC Counts Correlated Well for Fresh Samples

AO/PI and Methylene Blue TNC Counts Correlated Well for Fresh Samples



Sixteen samples (peripheral blood, cord blood, and bone marrow aspirates) were lysed / stained with 3% acetic acid / methylene blue and counted manually.

A second aliquot of each sample was stained with AO/PI without lysing and counted with the Cellometer Auto 2000.

Though the 3% acetic acid / methylene blue method provides accurate TNC counts, it does not measure cell viability.

The AO/PI method does not require cell lysis.

Trypan Blue and AO/PI Counts Correlated Well for Purified Samples

AO/PI and Methylene Blue TNC Counts Correlated Well for Fresh Samples


Excellent correlation between Trypan Blue and AO/PI for 24 purified samples (CD14+, CD19+, and CD34+ cells)

6 MNC samples did not correlate as well. Trypan Blue counts were higher due to RBC contamination. (red circles)

Trypan Blue or AO/PI can be used to accurately measure concentration and viability in isolated cell populations, such as CD34+, CD56+, and CD19+ cells.

Back to top ↑

Benefits of Cellometer Counting

Cellometer Vision and Auto 2000
  • Proven performance for samples containing red blood cells, platelets, and debris
  • One optimized dual-fluorescence method for analysis of both fresh and processed clinical samples, without lysing
  • Option to perform either traditional Trypan Blue or dual-fluorescence viability assays for purified samples
  • Demonstrated correlation to traditional counting methods
  • Improved accuracy and consistency of counting results
  • Automated saving and printing of results and images
  • Nexcelom Applications Specialists focus exclusively on cell counting and cell-based assays
  • Cellometer cell counters have been used to count more than 1,600 different cell types
  • Cellometer cell counters are used by the top ten pharmaceutical companies and all 40 NCI comprehensive cancer centers
  • Free Auto 2000 evaluation program
  • Free on-line and on-site training and customer support

Back to top ↑